Private working document · Jim Slagle · 28 Aug 2026 · est. 20–30 minutes
Companion signal source: forest.jslagle.net
(glance at the footer's version line first — reload if the browser has cached an old copy).
Supersedes the SENS value from the 27 Aug certification;
everything else from that run stands.
On 27 Aug the SENS trimmer was set against a misread level: the threshold landed at ≈0.79 V RMS (≈0 dBu — the very top of the RDL's −30…0 dBu trigger window), about 9 dB too high. Two reading traps compounded: a peak-to-peak figure was taken as an RMS figure (for a sine, Vpp is 2.83× the RMS), and the scope's channel probe-attenuation menu did not match the physical probe's 1×/10× switch. Only SENS is wrong. DELAY (14.5 s) is correct — do not touch it.
These are practical settings, not absolute engineering constants. The single aim: if the bench value is right, any technician can fine-tune the installed system entirely from the Mackie's knobs, without ever touching the trimmers again.
The value that achieves that: the relay just barely — but reliably — triggers at 0.28 V RMS, 1 kHz sine, measured at its input (screws 1–3). Why 0.28: it is ≈−9 dBu, near the middle of the RDL's window, and it is what the Mackie's AUX 1 send delivers with both aux knobs near 12 o'clock when channel 1 carries Bogdan's −12 dB speech reference — leaving the field technician turn-room in both directions. Anywhere in 0.25–0.30 V RMS is fine.
These three rows are the same signal. Aim for whichever row your instrument displays.
| Instrument & reading | Target | Accept |
|---|---|---|
| Siglent SDM350X DMM, AC volts, across screws 1–3 (preferred — it reads true RMS at 1 kHz and has no probe menu to betray you) | 0.28 V RMS | 0.25–0.30 V |
| Scope, automatic RMS / “Cyc RMS” measurement | 0.28 V | 0.25–0.30 V |
| Scope, peak-to-peak (counting divisions or auto Vpp) | 0.79 Vpp (0.40 V peak) | 0.71–0.85 Vpp |
The two traps, by name. ① 0.79 appears twice in this project's history. 0.79 Vpp is the correct target; 0.79 V RMS was the 27 Aug error — a 9 dB difference between two readings of the same digits. If your scope shows “0.79”, confirm which measurement it is displaying before believing anything. ② Probe menu: if a 10× probe is attached, the scope channel's probe setting must say 10× — otherwise every number is 10× off. Cross-check once: measure the same screws with the DMM; scope RMS and DMM must agree within a few percent. When they disagree, the DMM is the referee.
Identical rig to the 27 Aug certification (protocol): blue jumper bridging screws 2–3; signal hot → screw 1, signal ground → screw 3; 24 V supply on 4(+)/5(⏚); screws 6–12 empty. Workstation audio output wired to the signal leads; browser on the app. DMM (or scope) across screws 1–3. Trigger indication: the relay's audible click and the LED jumping from faint idle glow to bright — you know the two states by sight now.
SENS: just-trigger @ 0.28 V RMS (0.79 Vpp) 1 kHz · verified no-trigger @ −6 dB · 28 Aug 2026DELAY: 14.5 s — DO NOT ADJUST| Question | Answer |
|---|---|
| What voltage, and is it RMS or peak-to-peak? | 0.28 V RMS, which on a scope is 0.79 V peak-to-peak — same signal, two units. Prefer the DMM's AC-volts reading (true RMS, no probe trap). |
| Recommended way of getting that voltage? | App at MAX (0 dB) playing 1 kHz sine; trim Windows volume until the meter reads 0.28 V RMS; then freeze Windows volume and make all further changes in the app's exact-dB control. |
| Set SENS with 1 kHz sine or speech-band noise? | Set with sine, verify with speech. Sine is steady and instrument-readable; speech is the realistic spectrum. The sources are equal-RMS at the same Level, so one SENS setting serves both — if they ever disagree, suspect the rig, not the relay. |
| How far to back off for the “doesn't trigger here” test? | Exactly −6 dB — half the reference voltage. Field equivalent: CAL detent triggers, QUIET detent (CAL − 6) doesn't. |
| Observation | Result |
|---|---|
| Windows volume % at 0.28 V RMS reference | % |
| DMM at reference (app 0 dB, 1 kHz) | V RMS |
| Scope cross-check (if used): RMS / Vpp | |
| 5× re-trigger at reference | |
| 30 s at −6 dB (sine) — no trigger | |
| Speech-band: trigger at 0 / no-trigger at −6 | |
| Burst check: release time | s |
| Unit re-taped |
Prepared with AI assistance (Claude, Anthropic) from RDL's ST-ACR Series documentation (891-0031H), the 27–28 Aug bench record, and field SPL measurements of 28 Aug. For Jim's use; not for distribution.